Randomized evaluation of an inflatable femoral artery compression device after cardiac catheterization

5Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Mechanical femoral artery compression devices have several limitations. We compared a novel disposable belt-held pneumatic compression device to manual compression alone in 213 patients randomized into two equal groups. Both were comparable for age, gender, current therapy with aspirin (ASA) and warfarin, diameter of the arterial sheath, previous procedures via the same artery, procedure duration, and blood pressure. Manual compression time was 12 ± 3 minutes. Pneumatic compression was reduced during 60 minutes. Patient discomfort was assessed as none (82% vs 88%), mild (13% vs 8%), moderate (3% vs 4%), or severe (2% vs 0%) for the manual versus pneumatic group, respectively. Bleeding and hematoma occurred in 7.5% of patients with no difference between the treatment groups. However, manual compression was significantly more effective in the higher range of systolic blood pressure, and pneumatic in the lower range, with a cut point of approximately 170 mmHg. Predictors for bleeding were systolic blood pressure and dose of ASA. Among 113 patients with systolic blood pressure < 160 mmHg and low dose (75 mg) or no ASA, only 1 patient (0.9%) experienced bleeding while 31% of 16 patients with both elevated systolic blood pressure and high dose ASA (150-330 mg) bled. We conclude that pneumatic femoral artery compression does not reduce bleeding and hematoma compared with manual compression. The use of low dose (75 mg) or no ASA, as well as giving special attention to patients with elevated systolic blood pressure, may reduce the risk of bleeding after cardiac catheterization.

References Powered by Scopus

Complications of coronary arteriography from the Collaborative Study of Coronary Artery Surgery (CASS)

315Citations
N/AReaders
Get full text

Coronary arteriography 1984–1987: A report of the registry of the society for cardiac angiography and interventions. I. Results and complications

295Citations
N/AReaders
Get full text

Complication rate of coronary arteriography: a review of 5250 cases studied by a percutaneous femoral technique

206Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Early mobilization after protamine reversal of heparin following implantation of phosphorylcholine- coated stents in totally occluded coronary arteries

43Citations
N/AReaders
Get full text

Determining best practice: Comparison of three methods of femoral sheath removal after cardiac interventional procedures

32Citations
N/AReaders
Get full text

Effectiveness of mechanical compression devices in attaining hemostasis after femoral sheath removal

19Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Nordrehaug, J. E., Chronos, N. A. F., Priestley, K. A., Buller, N. P., Foran, J., Wainwright, R., … Sigwart, U. (1996). Randomized evaluation of an inflatable femoral artery compression device after cardiac catheterization. Journal of Interventional Cardiology, 9(5), 381–388. https://doi.org/10.1111/j.1540-8183.1996.tb00645.x

Readers' Seniority

Tooltip

Professor / Associate Prof. 1

50%

PhD / Post grad / Masters / Doc 1

50%

Readers' Discipline

Tooltip

Medicine and Dentistry 3

60%

Psychology 2

40%

Save time finding and organizing research with Mendeley