Effects of External Electrical and Magnetic Fields on Pacemakers and Defibrillators

  • Beinart R
  • Nazarian S
N/ACitations
Citations of this article
13Readers
Mendeley users who have this article in their library.

Abstract

The overall risk of clinically significant adverse events related to EMI in recipients of CIEDs is very low. Therefore, no special precautions are needed when household appliances are used. Environmental and industrial sources of EMI are relatively safe when the exposure time is limited and distance from the CIEDs is maximized. The risk of EMI-induced events is highest within the hospital environment. Physician awareness of the possible interactions and methods to minimize them is warranted.© 2013 American Heart Association, Inc.

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Beinart, R., & Nazarian, S. (2013). Effects of External Electrical and Magnetic Fields on Pacemakers and Defibrillators. Circulation, 128(25), 2799–2809. https://doi.org/10.1161/circulationaha.113.005697

Readers over time

‘17‘19‘20‘21‘22‘23‘2400.751.52.253

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 6

75%

Researcher 2

25%

Readers' Discipline

Tooltip

Materials Science 2

33%

Medicine and Dentistry 2

33%

Nursing and Health Professions 1

17%

Neuroscience 1

17%

Article Metrics

Tooltip
Mentions
Blog Mentions: 1
News Mentions: 1
References: 1
Social Media
Shares, Likes & Comments: 1

Save time finding and organizing research with Mendeley

Sign up for free
0