Syringe size and flow rate affect drug delivery from syringe pumps

64Citations
Citations of this article
61Readers
Mendeley users who have this article in their library.

Abstract

Purpose: To evaluate the effect of syringe size and infusion rate on drug delivery after vertical displacement of syringe pumps. Methods: Four syringes (10 ml, 20 ml, 30 ml, 50 ml) were studied at three infusion speeds (2 ml · hr-1, 1 ml · hr-1 and 0.5 ml · hr-1). Fluid delivery was measured gravimetrically using an electronic balance. Aspiration volume (amount of fluid retracted into the syringe-infusion line assembly) and zero-drug delivery time (time between the lowering of the syringe pump and reattainment of the initial weight) were determined after lowering the syringe pump 130 cm or 50 cm. Then, infusion bolus was measured after elevating the pump to its initial position. Syringe compliance was calculated from the occlusion release bolus at an occlusion pressure of 300 mmHg. Results: Zero-drug delivery time, aspiration volume and infusion bolus increased with syringe size (P < 0.0001). At 0.5 ml · hr-1, a zero-drug delivery time of 29.7 ± 0.6 min was recorded for the 50 ml syringe and 5.5 ± 0.15 min for the 10 ml syringe. As infusion rates increased, zero-drug delivery times decreased following a |/x ratio. Zero-drug delivery time highly correlated with calculated syringe compliance (Pearson's r = 0.97, P < 0.001), which was closely related to syringe plunger area (R2 = 0.93, P < 0.001). Conclusions: Vertical displacement of syringe pumps causes flow irregularities which correlate directly with syringe size and indirectly with infusion rate. Extreme vertical maneuvering of syringe pumps should be avoided infusion rates ≥ 1 ml · hr-1 and low-compliant, small syringes are recommended when highly concentrated drugs are administered.

References Powered by Scopus

Clinical events in association with timing of intraventricular hemorrhage in preterm infants

74Citations
N/AReaders
Get full text

Influence of infusion pump operation and flow rate on hemodynamic stability during epinephrine infusion

44Citations
N/AReaders
Get full text

Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors

43Citations
N/AReaders
Get full text

Cited by Powered by Scopus

The impact of carrier flow rate and infusion set dead-volume on the dynamics of intravenous drug delivery

78Citations
N/AReaders
Get full text

A Nonintrusive Flow Rate Sensor Based on Microwave Split-Ring Resonators and Thermal Modulation

61Citations
N/AReaders
Get full text

The delivery of drugs to patients by continuous intravenous infusion: Modeling predicts potential dose fluctuations depending on flow rates and infusion system dead volume

56Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Weiss, M., Hug, M. I., Neff, T., & Fischer, J. (2000). Syringe size and flow rate affect drug delivery from syringe pumps. Canadian Journal of Anesthesia, 47(10), 1031–1035. https://doi.org/10.1007/BF03024878

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 17

61%

Researcher 8

29%

Professor / Associate Prof. 2

7%

Lecturer / Post doc 1

4%

Readers' Discipline

Tooltip

Engineering 13

46%

Medicine and Dentistry 9

32%

Physics and Astronomy 4

14%

Materials Science 2

7%

Save time finding and organizing research with Mendeley

Sign up for free