A multicenter clinical evaluation of the Clot Signature Analyzer

24Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: The Clot Signature Analyzer (CSA) was designed to assess global hemostasis as a screening assay using non-anticoagulated whole blood. Three different measurements are produced by the instrument: platelet hemostasis time (PHT), clot time (CT), and collagen-induced thrombus formation (CITF). Objectives: The purpose of the present study was to determine normal ranges for these measurements and assess the performance of the CSA in patients with well-characterized hemostatic disorders and in normal subjects. Patients and methods: Four institutions participated in the study. Each established their own normal reference ranges. Patients with well-characterized hemostatic disorders and concurrent normal controls were subsequently examined. Results: Normal ranges between institutions were similar although statistically different. One hundred and eight patients were examined: 46 individuals with von Willebrand disease (VWD) (type 1, 26; type 2A, 11; type 2B, six; type 3, three); 38 patients with a coagulation factor deficiency; 13 individuals with platelet function defects; 10 patients taking warfarin; and one individual on low-molecular-weight heparin. Of these patients, 89% had at least one abnormality by CSA: 42/46 VWD patients, 35/38 coagulation protein defect patients, 9/13 patients with platelet function defects, 9/10 patients on warfarin and 1/1 patient on low-molecular-weight heparin. Of 116 normal subjects, 103 (89%) fell within the centers' normal range. These data suggest that the CSA has a good sensitivity for bleeding disorders. © 2004 International Society on Thrombosis and Haemostasis.

References Powered by Scopus

Aggregation of blood platelets by adenosine diphosphate and its reversal

3929Citations
N/AReaders
Get full text

A critical reappraisal of the bleeding time

633Citations
N/AReaders
Get full text

Description of an in vitro platelet function analyzer - PFA-100®

515Citations
N/AReaders
Get full text

Cited by Powered by Scopus

The obstetric and gynaecological management of women with inherited bleeding disorders-review with guidelines produced by a taskforce of UK Haemophilia Centre Doctors' Organization

239Citations
N/AReaders
Get full text

Hemostasis and menstruation: Appropriate investigation for underlying disorders of hemostasis in women with excessive menstrual bleeding

80Citations
N/AReaders
Get full text

Assays of different aspects of haemostasis - what do they measure?

61Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Fricke, W., Kouides, P., Kessler, C., Schmaier, A. H., Krijanovski, Y., Jagadeesan, K., & Joist, J. (2004). A multicenter clinical evaluation of the Clot Signature Analyzer. Journal of Thrombosis and Haemostasis, 2(5), 763–768. https://doi.org/10.1111/j.1538-7836.2004.00695.x

Readers' Seniority

Tooltip

Researcher 3

60%

PhD / Post grad / Masters / Doc 2

40%

Readers' Discipline

Tooltip

Medicine and Dentistry 4

57%

Chemical Engineering 1

14%

Nursing and Health Professions 1

14%

Arts and Humanities 1

14%

Save time finding and organizing research with Mendeley

Sign up for free