Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study

45Citations
Citations of this article
82Readers
Mendeley users who have this article in their library.

Abstract

Background: The MARVEL (Micra Atrial TRacking Using a Ventricular AccELerometer) 2 study assessed the efficacy of atrioventricular (AV) synchronous pacing with a Micra leadless pacemaker. Average atrioventricular synchrony (AVS) was 89.2%. Previously, low amplitude of the Micra-sensed atrial signal (A4) was observed to be a factor of low AVS. Objective: The purpose of this study was to identify predictors of A4 amplitude and high AVS. Methods: We analyzed 64 patients enrolled in MARVEL 2 who had visible P waves on electrocardiogram for assessing A4 amplitude and 40 patients with third-degree AV block for assessing AVS at rest. High AVS was defined as >90% correct atrial-triggered ventricular pacing. The association between clinical factors and echocardiographic parameters with A4 amplitude was investigated using a multivariable model with lasso variable selection. Variables associated with A4 amplitude together with premature ventricular contraction burden, sinus rate, and sinus rate variability (standard deviation of successive differences of P-P intervals [SDSD]) were assessed for association with AVS. Results: In univariate analysis, low A4 amplitude was inversely related to atrial function assessed by E/A ratio and eʹ/aʹ ratio, and was directly related to atrial contraction excursion (ACE) and atrial strain (Ɛa) on echocardiography (all P ≤.05). The multivariable lasso regression model found coronary artery bypass graft history, E/A ratio, ACE, and Ɛa were associated with low A4 amplitude. E/A ratio and SDSD were multivariable predictors of high AVS, with >90% probability if E/A <0.94 and SDSD <5 bpm. Conclusion: Clinical parameters and echocardiographic markers of atrial function are associated with A4 signal amplitude. High AVS can be predicted by E/A ratio <0.94 and low sinus rate variability at rest.

References Powered by Scopus

Regularization paths for generalized linear models via coordinate descent

12451Citations
N/AReaders
Get full text

An Overview of Heart Rate Variability Metrics and Norms

4289Citations
N/AReaders
Get full text

ROCR: Visualizing classifier performance in R

2486Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Predictors of time in target glucose range in real-world users of the MiniMed 780G system

75Citations
N/AReaders
Get full text

Leadless atrioventricular synchronous pacing in an outpatient setting: Early lessons learned on factors affecting atrioventricular synchrony

38Citations
N/AReaders
Get full text

Arrhythmias in Female Patients: Incidence, Presentation and Management

35Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Garweg, C., Khelae, S. K., Steinwender, C., Chan, J. Y. S., Ritter, P., Johansen, J. B., … Chinitz, L. (2020). Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study. Heart Rhythm, 17(12), 2037–2045. https://doi.org/10.1016/j.hrthm.2020.07.024

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 13

50%

Researcher 10

38%

Lecturer / Post doc 2

8%

Professor / Associate Prof. 1

4%

Readers' Discipline

Tooltip

Medicine and Dentistry 29

78%

Social Sciences 4

11%

Engineering 2

5%

Agricultural and Biological Sciences 2

5%

Save time finding and organizing research with Mendeley

Sign up for free