Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness

3Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Purpose: To compare readings of intraocular pressure (IOP) taken with the Goldmann applanation tonometer (GAT), the non-contact tonometer (NCT), and the rebound tonometer (RBT), and to compare their correlation with central corneal thickness (CCT). Methods: This was a prospective, cross-sectional, observational study to which patients above 18 years of age were enrolled. A total of 400 eyes of 200 non-glaucomatous patients underwent IOP recordings using the GAT, NCT, and RBT, and CCT was also noted. Informed consent of the patients was taken. The IOP readings taken via the three methods were compared and correlated with CCT. Paired t test was used to compare the two devices. Simple and multivariate linear regression analyses were used to study the relationship between factors. A P value less than 0.05 was considered significant. Correlation was determined using the Pearson correlation coefficient, and a Bland-Altman graph was plotted. Results: Mean IOP measured by the NCT was 15.65 ± 2.80 mmHg, by the RBT was 14.23 ± 3.05 mmHg, and by the GAT was 14.69 ± 2.97 mmHg. The mean CCT was 510.61 ± 33.83 microns. The difference between mean IOP recorded by the NCT and that by the RBT was 1.41 ± 2.39 mmHg, between the NCT and GAT was 0.95 ± 2.03 mmHg, and between the GAT and RBT was 0.45 ± 2.22 mmHg. The difference between the IOP values was statistically significant (P < 0.005). All tonometers showed a statistically significant correlation with CCT, but it was observed that the NCT had a stronger correlation (0.4037). Conclusion: The IOP readings taken by all the three methods were comparable; however, RBT values were closer to GAT values. CCT did influence the IOP values, and this should be kept in mind while evaluating.

References Powered by Scopus

Collagen fibrils appear more closely packed in the prepupillary cornea: Optical and biomechanical implications

183Citations
N/AReaders
Get full text

Assessment of True Intraocular Pressure: The Gap Between Theory and Practical Data

142Citations
N/AReaders
Get full text

The Influence of Corneal Properties on Rebound Tonometry

132Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Normal Range for the Schirmer Tear Test and Intraocular Pressure in Healthy Latvian Darkhead Lambs and Ewes

1Citations
N/AReaders
Get full text

Predicting 24-hour intraocular pressure peaks and averages with machine learning

0Citations
N/AReaders
Get full text

The efficacy and safety of anti-vascular endothelial growth factor combined with Ahmed glaucoma valve implantation in the treatment of neovascular glaucoma: a systematic review and meta-analysis

0Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Porwal, A. C., Shrishrimal, M., Punamia, R. P., & Mathew, B. C. (2023). Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness. Indian Journal of Ophthalmology, 71(5), 1927–1931. https://doi.org/10.4103/ijo.IJO_1982_22

Readers over time

‘23‘24‘25036912

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 5

71%

Professor / Associate Prof. 1

14%

Researcher 1

14%

Readers' Discipline

Tooltip

Medicine and Dentistry 4

57%

Nursing and Health Professions 1

14%

Biochemistry, Genetics and Molecular Bi... 1

14%

Engineering 1

14%

Article Metrics

Tooltip
Social Media
Shares, Likes & Comments: 6

Save time finding and organizing research with Mendeley

Sign up for free
0