Hounsfield unit recovery in clinical cone beam CT images of the thorax acquired for image guided radiation therapy

44Citations
Citations of this article
72Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A comprehensive artefact correction method for clinical cone beam CT (CBCT) images acquired for image guided radiation therapy (IGRT) on a commercial system is presented. The method is demonstrated to reduce artefacts and recover CT-like Hounsfield units (HU) in reconstructed CBCT images of five lung cancer patients. Projection image based artefact corrections of image lag, detector scatter, body scatter and beam hardening are described and applied to CBCT images of five lung cancer patients. Image quality is evaluated through visual appearance of the reconstructed images, HU-correspondence with the planning CT images, and total volume HU error. Artefacts are reduced and CT-like HUs are recovered in the artefact corrected CBCT images. Visual inspection confirms that artefacts are indeed suppressed by the proposed method, and the HU root mean square difference between reconstructed CBCTs and the reference CT images are reduced by 31% when using the artefact corrections compared to the standard clinical CBCT reconstruction. A versatile artefact correction method for clinical CBCT images acquired for IGRT has been developed. HU values are recovered in the corrected CBCT images. The proposed method relies on post processing of clinical projection images, and does not require patient specific optimisation. It is thus a powerful tool for image quality improvement of large numbers of CBCT images.

References Powered by Scopus

Practical cone-beam algorithm

5567Citations
N/AReaders
Get full text

Elastix: A toolbox for intensity-based medical image registration

3372Citations
N/AReaders
Get full text

Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization

1664Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Adaptive Radiotherapy for Anatomical Changes

189Citations
N/AReaders
Get full text

ScatterNet: A convolutional neural network for cone-beam CT intensity correction

128Citations
N/AReaders
Get full text

CBCT correction using a cycle-consistent generative adversarial network and unpaired training to enable photon and proton dose calculation

110Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Thing, R. S., Bernchou, U., Mainegra-Hing, E., Hansen, O., & Brink, C. (2016). Hounsfield unit recovery in clinical cone beam CT images of the thorax acquired for image guided radiation therapy. Physics in Medicine and Biology, 61(15), 5781–5802. https://doi.org/10.1088/0031-9155/61/15/5781

Readers over time

‘15‘16‘17‘18‘19‘20‘21‘22‘23‘2407142128

Readers' Seniority

Tooltip

Researcher 24

48%

PhD / Post grad / Masters / Doc 23

46%

Professor / Associate Prof. 2

4%

Lecturer / Post doc 1

2%

Readers' Discipline

Tooltip

Physics and Astronomy 21

43%

Medicine and Dentistry 14

29%

Engineering 10

20%

Nursing and Health Professions 4

8%

Save time finding and organizing research with Mendeley

Sign up for free
0