Real-time in vivo dosimetry for SBRT prostate treatment using plastic scintillating dosimetry embedded in a rectal balloon: A case study

12Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.

Abstract

A novel FDA approved in vivo dosimetry device system using plastic scintillating detectors placed in an endorectal balloon to provide real-time in vivo dosimetry for prostatic rectal interface was tested for use with stereotactic body radiotherapy (SBRT). The system was used for the first time ever to measure dose during linear accelerator based SBRT. A single patient was treated with a total dose of 36.25 Gy given in 5 fractions. Delivered dose was measured for each treatment with the detectors placed against the anterior rectal wall near the prostate rectal interface. Measured doses showed varying degrees of agreement with computed/ planned doses, with average combined dose found to be within 6% of the expected dose. The variance between measurements is most likely due to uncertainty of the detector location, as well as variation in the placement of a new balloon prior to each fraction. Distance to agreement for the detectors was generally found to be within a few millimeters, which also suggested that the differences in measured and calculated doses were due to positional uncertainty of the detectors during the SBRT, which had sharp dose falloff near the penumbra along the rectal wall. Overall, the use of a real time in vivo dosimeter provided a level of safety and improved confidence in treatment delivery. We are evaluating the device further in an IRB-approved prospective partial prostate SBRT trial, and believe further clinical investigations are warranted.

References Powered by Scopus

Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerations

428Citations
N/AReaders
Get full text

Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements

338Citations
N/AReaders
Get full text

In vivo dosimetry in external beam radiotherapy

277Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Estimating dose delivery accuracy in stereotactic body radiation therapy: A review of in-vivo measurement methods

41Citations
N/AReaders
Get full text

Real-time dosimetry with Yb-doped silica optical fibres

38Citations
N/AReaders
Get full text

Real-time dose-rate monitoring with gynecologic brachytherapy: Results of an initial clinical trial

22Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Cantley, J. L., Cheng, C. W., Jesseph, F. B., Podder, T. K., Colussi, V. C., Traughber, B. J., … Ellis, R. J. (2016). Real-time in vivo dosimetry for SBRT prostate treatment using plastic scintillating dosimetry embedded in a rectal balloon: A case study. Journal of Applied Clinical Medical Physics, 17(6), 305–311. https://doi.org/10.1120/jacmp.v17i6.6508

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 13

81%

Researcher 2

13%

Lecturer / Post doc 1

6%

Readers' Discipline

Tooltip

Medicine and Dentistry 5

42%

Physics and Astronomy 4

33%

Engineering 2

17%

Materials Science 1

8%

Save time finding and organizing research with Mendeley

Sign up for free