Evaluation of an automated reader and color interpretation-based immunoassays for multiplexed Drug-of-Abuse Testing in Urine

3Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

On-site drugs of abuse testing devices have undergone continuous improvement. We evaluated three devices with different designs: an automated reader, the Multi-Drug Screen Test Device with DxLINK (DxLINK; Innovacon, Alere, San Diego, USA) and two colorimetric immunoassays, the One Step Multi- Line Screen Panel with Integrated E-Z Split Key Cup II (E-Z Cup; Innovacon, Alere) and the One Step Multi-Drug Screen Panel card (Multi4 card; Alere, Abon Biopharm, Hangzhou, China). Eleven drugs [amphetamine, secobarbital, oxazepam, buprenorphine, benzoylecgonine, methylenedioxymethamphetamine (MDMA), 11-nor-9-carboxy-δ9-tetrahydrocannabinol (THC), methamphetamine, methadone, morphine and nortriptyline] were tested using the DxLINK and E-Z Cup. Four drugs (benzoylecgonine, THC, methamphetamine and morphine) were tested using the Multi4 card using control materials (Detectabuse Stat-Skreen; Biochemical Diagnostics, Edgewood, NY, USA). The concentrations (-50%, -25%, +25%, +50% and 3× cut-off values) of the control materials were confirmed by mass spectrometry. Concordance rates were calculated around cut-offs. All devices showed high overall agreement rates of >90% with a few exceptions: the DxLINK exhibited lower sensitivity for benzoylecgonine, methadone and nortriptyline (60% and 30%, 92% and 40%, and 96% and 60% sensitivity at +50% and +25% cut-off levels, respectively). The E-Z Cup exhibited lower sensitivity for oxazepam and nortriptyline (97% and 50%, and 97% and 40% sensitivity at +50% and +25% cut-off levels, respectively). We additionally evaluated test-band color by visual inspection using a standard color-scale card. When detailed color criteria for determination of positivity were applied for the E-Z Cup, using slightly less stringent criteria, oxazepam, buprenorphine, MDMA and nortriptyline showed increases in sensitivity from 70-80% to 90-100%, all with a specificity above 98%. Overall, all devices exhibited satisfactory performance at ±50% cut-off levels for commonly used drugs, with the exception of lower sensitivity for cocaine testing for DxLINK. Careful evaluation of devices and elaborate calibration of visual interpretation for determining positivitymay help improve the performance of these devices.

References Powered by Scopus

Urine drug screening: Practical guide for clinicians

445Citations
N/AReaders
Get full text

Tetrahydrocannabinol stability in whole blood: Plastic versus glass containers

104Citations
N/AReaders
Get full text

A comparison between on-site immunoassay drug-testing devices and laboratory results

66Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Assessing the reliability of commercially available point of care in various clinical fields

10Citations
N/AReaders
Get full text

A Membrane-Modulated Centrifugal Microdevice for Enzyme-Linked Immunosorbent Assay-Based Detection of Illicit and Misused Drugs

8Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Kim, S. Y., Kim, H., Park, Y., Lim, J., Kim, J., Koo, S. H., & Kwon, G. C. (2017). Evaluation of an automated reader and color interpretation-based immunoassays for multiplexed Drug-of-Abuse Testing in Urine. Journal of Analytical Toxicology, 41(5), 412–420. https://doi.org/10.1093/jat/bkx014

Readers over time

‘17‘18‘19‘20‘21‘22‘23‘240481216

Readers' Seniority

Tooltip

Researcher 7

54%

PhD / Post grad / Masters / Doc 5

38%

Professor / Associate Prof. 1

8%

Readers' Discipline

Tooltip

Pharmacology, Toxicology and Pharmaceut... 4

29%

Chemistry 4

29%

Medicine and Dentistry 4

29%

Biochemistry, Genetics and Molecular Bi... 2

14%

Save time finding and organizing research with Mendeley

Sign up for free
0