Photon counting energy dispersive detector arrays for X-ray imaging

268Citations
Citations of this article
103Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The development of an innovative detector technology for photon-counting in X-ray imaging is reported. This new generation of detectors, based on pixellated cadmium telluride (CdTe) and cadmium zinc telluride (CZT) detector arrays electrically connected to application specific integrated circuits (ASICs) for readout, will produce fast and highly efficient photon-counting and energy-dispersive X-ray imaging. There are a number of applications that can greatly benefit from these novel imagers including mammography, planar radiography, and computed tomography (CT). Systems based on this new detector technology can provide compositional analysis of tissue through spectroscopic X-ray imaging, significantly improve overall image quality, and may significantly reduce X-ray dose to the patient. A very high X-ray flux is utilized in many of these applications. For example, CT scanners can produce ∼ 100 Mphotons mm2 /s in the unattenuated beam. High flux is required in order to collect sufficient photon statistics in the measurement of the transmitted flux (attenuated beam) during the very short time frame of a CT scan. This high count rate combined with a need for high detection efficiency requires the development of detector structures that can provide a response signal much faster than the transit time of carriers over the whole detector thickness. We have developed CdTe and CZT detector array structures which are 3 mm thick with 16 ×16 pixels and a 1 mm pixel pitch. These structures, in the two different implementations presented here, utilize either a small pixel effect or a drift phenomenon. An energy resolution of 4.75% at 122 keV has been obtained with a 30 ns peaking time using discrete electronics and a57 Co source. An output rate of 6 × 10 6 counts per second per individual pixel has been obtained with our ASIC readout electronics and a clinical CT X-ray tube. Additionally, the first clinical CT images, taken with several of our prototype photon-counting and energy-dispersive detector modules, are shown. © 2006 IEEE.

References Powered by Scopus

K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors

515Citations
N/AReaders
Get full text

Unipolar Charge Sensing with Coplanar Electrodes—Application to Semiconductor Detectors

283Citations
N/AReaders
Get full text

Beam hardening artefacts in computed tomography with photon counting, charge integrating and energy weighting detectors: A simulation study

105Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Photon-counting CT: Technical principles and clinical prospects

862Citations
N/AReaders
Get full text

CT artifacts: Causes and reduction techniques

662Citations
N/AReaders
Get full text

The evolution of image reconstruction for CT—from filtered back projection to artificial intelligence

402Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Iwanczyk, J. S., Nygård, E., Meirav, O., Arenson, J., Barber, W. C., Hartsough, N. E., … Wessel, J. C. (2009). Photon counting energy dispersive detector arrays for X-ray imaging. IEEE Transactions on Nuclear Science, 56(3), 535–542. https://doi.org/10.1109/TNS.2009.2013709

Readers over time

‘10‘11‘12‘13‘14‘15‘16‘17‘18‘19‘20‘21‘22‘23‘240481216

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 40

51%

Researcher 31

40%

Professor / Associate Prof. 5

6%

Lecturer / Post doc 2

3%

Readers' Discipline

Tooltip

Engineering 35

51%

Physics and Astronomy 24

35%

Medicine and Dentistry 8

12%

Computer Science 2

3%

Article Metrics

Tooltip
Mentions
References: 2

Save time finding and organizing research with Mendeley

Sign up for free
0