Abstract

An electrochemical quartz crystal microbalance (EC‐QCM) is a versatile gravimetric technique that allows for parallel characterization of mass deposition and electrochemical properties. Despite its broad applicability, simultaneous characterization of two electrodes remains challenging due to practical difficulties posed by the dampening from fixture parasitics and the dissipative medium. In this study, we present Fa dual electrochemical QCM (dual EC‐QCM) that is employed in a three‐electrode configuration to enable consequent monitoring of mass deposition and viscous loading on two crystals, the working electrode (WE) and the counter electrode (CE). A novel correction approach, along with a three standard complex impedance calibration, is employed to overcome the effect of dampening while keeping high spectral sensitivity. Separation of viscous loading and rigid mass deposition is achieved by robust characterization of the complex impedance at the resonance frequency. Validation of the presented system is done by cyclic voltammetry characterization of Ag underpotential deposition on gold. The results indicate mass deposition of 412.2 ng for the WE and 345.6 ng for the CE, reflecting a difference of the initially‐present Ag adhered to the surface. We also performed higher harmonic measurements that further corroborate the sensitivity and reproducibility of the dual EC‐QCM. The demonstrated approach is especially intriguing for electrochemical energy storage applications where mass detection with multiple electrodes is desired.

References Powered by Scopus

Verwendung von Schwingquarzen zur Wägung dünner Schichten und zur Mikrowägung

9733Citations
N/AReaders
Get full text

The oscillation frequency of a quartz resonator in contact with liquid

1170Citations
N/AReaders
Get full text

Characterization of a Quartz Crystal Microbalance with Simultaneous Mass and Liquid Loading

958Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Practical Use of Quartz Crystal Microbalance Monitoring in Cartilage Tissue Engineering

8Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Tóth, D., Kasper, M., Alic, I., Awadein, M., Ebner, A., Baney, D., … Kienberger, F. (2021). High‐sensitivity dual electrochemical qcm for reliable three‐electrode measurements. Sensors, 21(8). https://doi.org/10.3390/s21082592

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 4

57%

Lecturer / Post doc 2

29%

Researcher 1

14%

Readers' Discipline

Tooltip

Chemistry 2

33%

Engineering 2

33%

Chemical Engineering 1

17%

Materials Science 1

17%

Article Metrics

Tooltip
Mentions
Blog Mentions: 1

Save time finding and organizing research with Mendeley

Sign up for free