Comparison of the accuracy of noninvasive hemoglobin sensor (NBM-200) and portable hemoglobinometer (hemocue) with an automated hematology analyzer (LH500) in blood donor screening

49Citations
Citations of this article
73Readers
Mendeley users who have this article in their library.

Abstract

Background: The Hb levels of prospective blood donors are usually determined using a finger prick test. A new noninvasive Hb device has the advantage of not causing any sampling pain. The purpose of this study was to evaluate the accuracy of the noninvasive Hb sensor and to compare its measurements with those of a currently used portable hemoglobinometer. Methods: Hb was measured using a noninvasive Hb sensor (NBM-200; OrSense, Israel), a portable hemoglobinometer (HemoCue; HemoCue AB, Sweden), and an automated hematology analyzer (LH500; Beckman Coulter, USA). The correlations between Hb measurements taken by the NBM-200 and HemoCue with those by an automated hematology analyzer were assessed using intraclass correlation coefficients (ICCs). Hb measurements were compared among 3 different Hb level groups. Results: The mean Hb values of 506 blood donors were 14.1 g/dL by the NBM-200, 14.0 g/dL by the LH500, and 14.3 g/dL by the HemoCue. The correlation between the LH500 and the NBM-200 was substantial (ICC=0.69), while that between the LH500 and the HemoCue agreed almost perfectly (ICC=0.86). Conclusions: The possibility to judge to be eligible for donors who are ineligible to donate was substantial when using NBM-200. Even though the NBM-200 has the apparent advantage of noninvasiveness, its use in pre-screening should be given meticulous attention. Since pre-donation testing is crucial to protecting donors' health, complete evaluation of the instrument should be performed prior to use. © The Korean Society for Laboratory Medicine.

References Powered by Scopus

The measurement of observer agreement for categorical data

60246Citations
N/AReaders
Get full text

Hemoglobin measured by Hemocue and a reference method in venous and capillary blood: A validation study

124Citations
N/AReaders
Get full text

Non-invasive measurement of hemoglobin: Assessment of two different point-of-care technologies

91Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings

119Citations
N/AReaders
Get full text

Systematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7™ or Pronto-7™) and HemoCue® absorption spectrometers (B-Hemoglobin or 201+) with laboratory haemoglobin estimation

86Citations
N/AReaders
Get full text

Accuracy of continuous noninvasive hemoglobin monitoring: A systematic review and meta-analysis

85Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Kim, M. J., Park, Q., Kim, M. H., Shin, J. W., & Kim, H. O. (2013). Comparison of the accuracy of noninvasive hemoglobin sensor (NBM-200) and portable hemoglobinometer (hemocue) with an automated hematology analyzer (LH500) in blood donor screening. Annals of Laboratory Medicine, 33(4), 261–267. https://doi.org/10.3343/alm.2013.33.4.261

Readers over time

‘13‘15‘16‘17‘18‘19‘20‘21‘22‘23‘24‘250481216

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 25

63%

Researcher 12

30%

Professor / Associate Prof. 2

5%

Lecturer / Post doc 1

3%

Readers' Discipline

Tooltip

Medicine and Dentistry 22

61%

Engineering 6

17%

Agricultural and Biological Sciences 5

14%

Nursing and Health Professions 3

8%

Save time finding and organizing research with Mendeley

Sign up for free
0